Principles of three different automated urine microscopy analysers - Sysmex UF1000i, Menarini sediMAX and Iris iQ200
The
use of automated system in urine analysis reduces the time and cost used in the
manual routine analysis. In addition, automated system analyses a high number
of samples in a short time (up to 100 samples per hour). There are a few
methods available for automated system which includes the Sysmex UF1000i, Menarini sediMAX and Iris iQ200.
The Sysmex UF1000i method uses the urine flow cytometry
technology to detect red blood cells, white blood cells, bacteria, epithelial
cells and casts in urine. The automated system dilutes 200µL of the urine
sample with citrate buffer solution and cationic surfactant with an addition of
40µL of fluorescent dye to complete the mixture. This mixture is then passed
through a sheath of flow which is exposed to a laser beam to detect the
particles in the stained urine sample with the fluorescent dye staining the nucleic acid in
bacteria. The identification and characterisation of particles are based on the
fluorescence intensity, forward scatter light and side scatter light intensity
which generates electrical signal patterns by photomultiplier conversion which
is then measured.
The Menarini sediMAX
is an automated microscopy image based sediment analyser used in urine
microscopy. The analyser produces a monolayer of urine sediment by
centrifugation in a special cuvette. The sediment is analyzed by a bright field
microscope and digital camera to capture and categorize 15 particle images
based upon size and shape using image processing software. The digital images
can be viewed by a trained laboratory staff with zoom capability. The instrument has a high sensitivity (>80%)
in the detection of RBC and WBC but has a low sensitivity (approximately 50%)
when compared with manual microscopy.
The
Iris iQ200 method uses digital imaging and auto particle recognition to classify
and quantify urine particles in uncentrifuged urine based upon size and shape. It
classifies images such as WBC, RBC, bacteria, casts, yeast and crystals into
categories. The digital images generated are collected, approximately 500
frames per sample, then viewed and classified by a trained laboratory staff.
This is usually achieved without the use of manual microscopy.
The
advantages and disadvantages of the automated urine systems
(Sysmex UF1000i, SediMAX and Iris iQ200) are listed below.
Analyser
|
Advantages
|
Disadvantages
|
Sysmex UF1000i
|
Uses uncentrifuged
urine samples.
Improved bacteria
counting and flags for UTI.
Urgent urine samples
can be processed by operating the analyser manually.
Throughput of 100
samples per hour.
|
Cannot differentiate
clinically significant pathologic casts cell from benign casts.
Manual microscopy is
carried out on flagged results.
|
sediMAX
|
This analyser does not
require any reagents.
Requires minimal
laboratory technician training to interpret images that are similar to manual
microscopy.
Results may be
available for viewing from the stored digital images
Throughput of 80
samples per hour.
|
Yeast is mistaken for
RBCs.
Epithelial cells are
mistaken for WBCs.
Particle loss may
occur due to the ultracentrifugation of urine samples.
|
Iris iQ200
|
Uses uncentrifuged
urine samples.
Accurately quantifies
bacteria.
Results may be
available for viewing from the stored digital images
Throughput of up to
101 samples per hour.
|
Crystals are mistaken
for RBCs.
There is no sufficient
resolution in order to classify all dysmorphic RBC variants which prompt
manual microscopy.
Requires considerable
laboratory technician training for viewing digital images that are different
from manual microscopy.
|
Detection of biochemical markers and their significance in urine samples
Biochemical markers in urine such as glucose, bilirubin,
nitrite, ketone, specific gravity, blood, pH, protein, leukocytes and
urobilinogen can be used for UTI. It can be done conveniently as a point of
care test or near patients test outside the laboratory. This can be done manually
(dipstick) or automated (Siemens Clinitek and Roche Urisys 1100).
A urine dipstick Bayer urinalysis strip such as the
Multistix 10 SG is a narrow plastic strip with squares of different colours on
it which represents each of the biochemical markers present. The strip is usually
dipped in the urine sample to be tested and colour changes noted within 1
minute. The results are then interpreted as shown in the table below.
Table 2: Biochemical markers and its clinical relevance
Biochemical
marker
|
Clinical
relevance
|
Nitrite
|
Bacteria presence in urine
|
Leucocyte esterase
|
WBCs presence in urine
|
Urobilinogen
|
Possible liver disease
|
Bilirubin
|
Possible liver disease or RBCs breakdown
|
pH
|
Acidity of the urine sample and useful in interpretation
of microscopic crystals
|
Ketone
|
Product of fat metabolism
|
Specific gravity
|
Concentration of urine
|
Glucose
|
Diabetes mellitus
|
Protein
|
Glomerular disease
|
Hemaglobin
|
RBC trauma or enzyme deficiencies
|
Automated
systems such as Siemens Clinitek analyser is an easy to use chemistry analyser
that is used to read Bayer urinalysis strip such as the Multistix 10 SG. The
dipped urinalysis strip is inserted into the analyser and result reported
automatically.
The Roche
Urisys 1100 urine analyser is a semi automated analyser used for in vitro semi
quantitative determination of urine analyte. It is a reflectance photometer designed
to read and evaluate the results of urine test strips such as Chemstrip 10MD,
Chemstrip 7 and Chemstrip 5 OB. The test strip is placed on the sliding tray and
a stepping motor moves it under the reading head which remains stationary. The
analyser then reads the reference pad followed by each of the pads on the
strip. The reading head contains LEDs that emit light at various wavelengths. The
LED emits light of a defined wavelength onto the surface of the test pad at an
optimum angle. The light hitting the test zone is reflected proportionally to
the colour produced on the test pad and is picked up by the detector, a
phototransistor positioned directly above the test zone. The phototransistor
sends an analog electrical signal to an A/D converter, which changes it to
digital form. The microprocessor then converts this digital reading to a
relative reflectance value by referring it to a calibration standard. Finally,
the system compares the reflectance value with the defined range limits (reflectance
values which are programmed into the analyzer for each parameter) and outputs a
semi-quantitative result. Each test pad is read photometrically after about
55-65 seconds. In strongly alkaline urine samples, the Urisys 1100 Urine
Analyzer automatically corrects the result of the specific gravity test.
No comments:
Post a Comment